Ultrasound Attenuation: Cutoff Values, Specificity, and Sensitivity

Head-to-head diagnostic thresholds across five commercial ultrasound platforms benchmarked against MRI-PDFF as the reference standard.

What This Means for Patients

Different hospitals use different ultrasound machines, and each brand reports liver fat in its own unit of measurement. A reading of 0.68 on a GE UGAP system does not use the same scale as 288 on an Echosens FibroScan CAP report. This table lets you and your clinician match the exact brand on your imaging report to the cutoffs validated against MRI-PDFF.

Diagnostic Performance of Ultrasound Attenuation Using MRI-PDFF as the Reference Standard1

Manufacturer Technique (Validation Population) S0 vs S1–S3 S0–S1 vs S2–S3 S0–S2 vs S3
Siemens2 BSC, 1/cm-sr (MASLD) 0.0038 — —
Sensitivitya (%) 93 — —
Specificityb (%) 97 — —
Echosens3 CAP™, dB/m (MASLD) 288 — —
Sensitivitya (%) 75 — —
Specificityb (%) 77 — —
Canon Medical4 ATI, dB/cm/MHz (MASLD) 0.63 0.72 —
Sensitivitya (%) 80 100 —
Specificityb (%) 89 78 —
GE Healthcare5 UGAP, dB/cm/MHz (MASLD) 0.63 0.68 0.74
Sensitivitya (%) 80 83 87
Specificityb (%) 91 74 79
SuperSonic Imagine6 SSE, mm/μs (not specified) 1.537 1.511 —
Sensitivitya (%) 80 100 —
Specificityb (%) 86 96 —