Ultrasound Attenuation: Cutoff Values, Specificity, and Sensitivity
Head-to-head diagnostic thresholds across five commercial ultrasound platforms benchmarked against MRI-PDFF as the reference standard.
Different hospitals use different ultrasound machines, and each brand reports liver fat in its own unit of measurement. A reading of 0.68 on a GE UGAP system does not use the same scale as 288 on an Echosens FibroScan CAP report. This table lets you and your clinician match the exact brand on your imaging report to the cutoffs validated against MRI-PDFF.
Diagnostic Performance of Ultrasound Attenuation Using MRI-PDFF as the Reference Standard1
| Manufacturer | Technique (Validation Population) | S0 vs S1–S3 | S0–S1 vs S2–S3 | S0–S2 vs S3 |
|---|---|---|---|---|
| Siemens2 | BSC, 1/cm-sr (MASLD) | 0.0038 | — | — |
| Sensitivitya (%) | 93 | — | — | |
| Specificityb (%) | 97 | — | — | |
| Echosens3 | CAP™, dB/m (MASLD) | 288 | — | — |
| Sensitivitya (%) | 75 | — | — | |
| Specificityb (%) | 77 | — | — | |
| Canon Medical4 | ATI, dB/cm/MHz (MASLD) | 0.63 | 0.72 | — |
| Sensitivitya (%) | 80 | 100 | — | |
| Specificityb (%) | 89 | 78 | — | |
| GE Healthcare5 | UGAP, dB/cm/MHz (MASLD) | 0.63 | 0.68 | 0.74 |
| Sensitivitya (%) | 80 | 83 | 87 | |
| Specificityb (%) | 91 | 74 | 79 | |
| SuperSonic Imagine6 | SSE, mm/μs (not specified) | 1.537 | 1.511 | — |
| Sensitivitya (%) | 80 | 100 | — | |
| Specificityb (%) | 86 | 96 | — |