Ultrasound-Based TestingFibrosis Assessment
Imaging-Based Diagnostic Tests • Diagnostic Tests With Ultrasound

Differences in Transient or Shear Wave Elastography Technologies

Comparison of stimulus frequency, image guidance, sampled tissue volume (ROI size), and probe configurations across 1D TE, Visual TE, 2D TE, 3D S-WAVE, pSWE, 2D SWE, and MRE.

What This Means for Patients

Not all stiffness scans sample the same amount of liver tissue. A standard 1D FibroScan measures a cylinder of about 3 cubic centimeters, while newer 2D, 3D ultrasound, and MR elastography systems examine 20 to 250 cubic centimeters of liver. Sampling a larger slice of liver reduces the odds of getting a misleading score if scarring is patchy from one spot to another.

Modality & System Specifications Across Elastography Platforms

Modality Example System Frequency,a Hz Measurement Location ROI Size # of Probes
1D TE1,2 Echosens FibroScan®b 50 Variable, probe-dependent 3 cubic cm Up to 3c
Visual TE3,4 Mindray Hepatus 50 Image-guided ≥6 cubic cm Up to 2
2D TE5 E-Scopics Hepatoscope 50 Image-guided 22.5 cubic cm 1
3D S-WAVE6,7 Sonic Incytes Velacur™ 40–70 AI-guided 100 cubic cm 1
Point SWE8 Philips ElastQ 100–500 Image-guided 0.5–1 cubic cm Varies by system
2D SWE8 GE LOGIQ™ E10 100–500 Image-guided 20 cubic cm Varies by system
MRE8 GE MR Touch 60 User-defined 250 cubic cm NA