Differences in Transient or Shear Wave Elastography Technologies
Comparison of stimulus frequency, image guidance, sampled tissue volume (ROI size), and probe configurations across 1D TE, Visual TE, 2D TE, 3D S-WAVE, pSWE, 2D SWE, and MRE.
Not all stiffness scans sample the same amount of liver tissue. A standard 1D FibroScan measures a cylinder of about 3 cubic centimeters, while newer 2D, 3D ultrasound, and MR elastography systems examine 20 to 250 cubic centimeters of liver. Sampling a larger slice of liver reduces the odds of getting a misleading score if scarring is patchy from one spot to another.
Modality & System Specifications Across Elastography Platforms
| Modality | Example System | Frequency,a Hz | Measurement Location | ROI Size | # of Probes |
|---|---|---|---|---|---|
| 1D TE1,2 | Echosens FibroScan®b | 50 | Variable, probe-dependent | 3 cubic cm | Up to 3c |
| Visual TE3,4 | Mindray Hepatus | 50 | Image-guided | ≥6 cubic cm | Up to 2 |
| 2D TE5 | E-Scopics Hepatoscope | 50 | Image-guided | 22.5 cubic cm | 1 |
| 3D S-WAVE6,7 | Sonic Incytes Velacur™ | 40–70 | AI-guided | 100 cubic cm | 1 |
| Point SWE8 | Philips ElastQ | 100–500 | Image-guided | 0.5–1 cubic cm | Varies by system |
| 2D SWE8 | GE LOGIQ™ E10 | 100–500 | Image-guided | 20 cubic cm | Varies by system |
| MRE8 | GE MR Touch | 60 | User-defined | 250 cubic cm | NA |