Ultrasound-Based TestingSteatosis Assessment
Imaging-Based Diagnostic Tests • Diagnostic Tests With Ultrasound

Ultrasound Attenuation

Quantitative measurement of ultrasound signal loss through hepatic tissue to estimate liver steatosis alongside elastography.

What This Means for Patients

If you have had a FibroScan, you usually get two numbers back: a stiffness score in kPa (scarring) and a CAP score in dB/m (liver fat). Fatty liver tissue absorbs sound waves faster than lean liver tissue, and ultrasound attenuation measures that exact drop-off. A CAP reading at or above 288 dB/m or a UGAP reading above 0.63 dB/cm/MHz points to elevated liver fat, giving you a concrete number you can track over time as you work on nutrition, weight, or MASH treatment.

What Is Ultrasound Attenuation?

  • Attenuation is a quantitative ultrasound technique that measures the reduction in ultrasound signal amplitude as it passes through liver tissue, providing an estimate of hepatic steatosis1,2

How Is It Performed?

  • Attenuation is measured by analyzing the attenuation of ultrasound waves as they pass through liver tissue2
  • As liver fat (steatosis) increases, ultrasound waves are more strongly attenuated1

Use in MASLD/MASH Care

  • Clinical guidelines recommend quantitative liver fat assessment (eg, CAP™ by FibroScan®) in the broader evaluation of MASLD/MASH3,4
  • Attenuation (eg, CAP™) can be measured as part of an elastography exam (eg, VCTE™) for dual assessment of steatosis and fibrosis to support MASLD risk stratification and guide treatment decisions5

Strengths

  • Noninvasive quantifiable estimate of liver fat1
  • Good diagnostic accuracy with established cutoffs, sensitivity, and specificity1,6
  • Attenuation (eg, CAP™) has been used to monitor change in liver fat in a MASH trial7

Limitations

  • Measurement scale/output can vary across the various imaging platforms8
  • Accuracy can be affected by technical factors such as transducer frequency and gain settings9

Published Cutoff Values Suggestive of Elevated Liver Fat

Parameter Cutoff Sensitivity,a,b % Specificity,a,c %
CAP™3,4 ≥288 dB/m 75% 77%
UGAP4 >0.63 dB/cm/MHz 80% 91%